IVF Laminar Air Flow

Home – IVF Laminar Air Flow

Laminar Air Flow Manufacturers in Chennai – IVF Laminar Air Flow

Are you looking for cutting-edge technologies for your IVF lab? There is no need to look any further! We are the pioneers of IVF laminar airflow solutions, and we are committed to revolutionising the assisted reproductive technology business. We are your trusted partner in establishing the ideal atmosphere for successful IVF operations, with a focus to innovation and precision.

IVF laminar airflow is a special form of laminar airflow cabinet used in in vitro fertilisation (IVF) facilities. It creates a regulated atmosphere in which pollutants are kept out of the work area, which is critical for the effectiveness of IVF treatments.

HEPA filters are used to remove particles and bacteria from the air in an IVF laminar airflow cabinet. Additionally, the cabinet features a positive pressure differential, which means that the air inside the cabinet is always under pressure, preventing pollutants from entering from the outside.

Enquire for more details : [email protected]

Product “Pureair” Vertical laminar Airflow – IVF Version
Application In- Vitro Fertilization Application
Dimensions As per drawings enclosed.
MOC Galvanized Iron with PU coated / Stainless Steel – Satin Finish / Wooden laminated
Pure Filter “Pureair” Make Pre Filter with Micro Glass Fiber Media of EU – 4 Rating.
Supply HEPA Filter Imported Minipleat HEPA Filter efficiency 99.97% Down to 0.3µ H-13 Rating.
Mounting Unique Spring Suspension Systems to take care of Vibration and Noise.
Working Table S.S. 304 with Stabilized Heating Table with Microscope Arrange ment
Side Panel Accessory No. Clear Glass Side Panels / Clear Acrylic panel / UV Protected Polycarbonate panel
Pressure Gauge DWYER, USA Minihelix Differential Pressure Gauge / “Pureair” Inclined manometer.
Lights Fluorescent Tube with Milky White Diffuser.
Other Electricals 3 Pin Socket, 15 AMP – 1 No. – On / Off Switches with Indicator.
Air Cleanliness Class ISO – 5 (as per ISO – 14644-1).
Air Flow rate 0.45 ± 0.05mps. or 90 FPM ± 10 FPM
Illumination Of Light Minimum 800 Lux on work table
Noise Level Less than 67 db on Scale ‘A’ When Ambient sound Level is not Greater than 57 db on Scale ‘A’
Vibration Level Minimum

Fume Exhaust Hood Manufacturers in Chennai

Biosafety Cabinets Manufacturers in Chennai

Bypass Type Fume Exhaust Hood Manufacturers in Chennai