Horizontal Airflow
Home – Horizontal Airflow
Laminar Air Flow Manufacturers in Chennai – Horizontal Laminar Air Flow
Horizontal airflow is produced by a fan that blows air from the back of the cabinet towards the front. The air moves horizontally, perpendicular to the work surface. Horizontal airflow cabinets are often employed for operations requiring a higher level of protection, such as handling hazardous materials or biological samples.
Horizontal airflow cabinets are intended to offer a controlled environment in which pollutants are kept out of the work area. HEPA filters are used to remove particles and bacteria from the air. Additionally, the cabinets contain a positive pressure differential, which means that the air within the cabinet is constantly under pressure, preventing pollutants from entering from the outside.
Enquire for more details : [email protected]
Product | “Pureair” Horizontal laminar Airflow – Standard Version |
Application | Tissue Culture / Biotechnology |
Dimensions | As per drawings enclosed. |
MOC | Galvanized Iron with PU coated / Stainless Steel – Satin Finish / Wooden laminated (Except Filters and Motor – Blower Filters). |
Pre-Filter | “Pureair” Make Pre-Filter with Micro Glass Fibre Media of EU – 4 Rating. |
Supply HEPA Filter | Imported Miniplate HEPA Filter efficiency 99.97% Down to 0.3µ H-13 Rating. |
Motor Blower Assembly | Induction Motor with Centrifugal Blower Statically and Dynamically balanced with Suitable Rating and Size. |
Mounting | Unique Spring Suspension Systems to take care of Vibration and Noise. |
Working Table | S.S. 304 with Due Reinforcement for Better Performance and Stability. |
Side Panel Accessory No. | Clear Glass Side Panels / Clear Acrylic panel / UV Protected Polycarbonate panel |
Pressure Gauge | DWYER, USA Minihelix Differential Pressure Gauge / “Pureair” Inclined manometer. |
Lights | Fluorescent Tube with Milky White Diffuser. |
Other Electricals | 3 Pin Socket, 15 AMP – 1 No. – On / Off Switches with Indicator. |
Air Cleanliness | Class ISO – 5 (as per ISO – 14644-1). |
Air Flow rate | 0.45 ± 0.05mps. or 90 FPM ± 10 FPM |
Illumination Of Light | Minimum 800 Lux on work table |
Noise Level | Less than 67 db on Scale ‘A’ When Ambient sound Level is not Greater than 57 db on Scale ‘A’ |
Vibration Level | Minimum |