Class II Type B2 Cabinets
Home -Class II Type B2
Class II Type B2 Biosafety Cabinet Manufacturers in Chennai
Among Class II cabinets, Class II Type B2 cabinets provide the maximum level of protection. With a unique circulation pattern that draws air downward across the work surface and exhausts it via HEPA filters, they safeguard both employees and products. This “total exhaust” design guarantees that all of the air within the cabinet is securely filtered and evacuated. Class II Type B2 cabinets are appropriate for handling dangerous biological agents and volatile chemicals.
Enquire for more details : [email protected]
Working Area – Stainless Steel IS 304 Grade Satin Finish
Product | Bio safety Cabinet – Class II Type B2 |
Air Balancing | 0% recirculation.100% exhaust |
Protection | Product & personnel & environmental protection |
Bio safety Level | Bio safety Level 1 , 2 & 3 |
Exhaust Duct Requirement | Hard Duct Only |
Control Plenum surrounded by | Negative Plenum |
Cleanliness level | ISO class 5 as per ISO 14644-1 |
Average Airflow | Inflow Velocity – Minimum 0.45 m/s ( 90 FPM) at 8” front opening |
Velocity | Down Velocity – 0.30 m/s ( 65 FPM) |
Noise level | 60 db on A scale |
Light Intensity | Minimum 900 lux inside the working area at ambient zero |
MOC Main Body – | Galvanized Iron with PU coated / Stainless Steel – Satin Finish |
Working Area – | Stainless Steel IS 304 Grade Satin Finish |
Supply Filter | HEPA Filter having EU 13 rating with suitable size. |
Efficiency | 99.99% down to 0.3 Microns. |
Exhaust Filter | HEPA Filter having EU 13 rating with suitable size. |
Efficiency | 99.99% down to 0.3 Microns. |
Pre Filter | “Pureair” make, EU 4 rating with suitable size. |
Efficiency | 90% down to 10 Microns. |
Noise Level | Less than 70 db ‘A’. |
Vibration Level Instrumentation | Minimum. |
Differential Pressure Gauge | Magnehelic Gauge – 01 No. (make : Dwyer, USA) with each unit (To measure differential pressure across the supply HEPA filter). |
Motor Blower Assembly | Statically & Dynamically balanced with suitable rating and size. |
Supply & Exhaust | (Induction motor with centrifugal blower) A Mounting unique spring suspension system to take care of vibration and noise. |
3 Pins Switch & Socket | 15 AMPS – 1 set in each unit. |
Other Switches | On/Off switches for motor blower & CFL lights. |
Microprocessor Control | Airflow display and UV Interlocking and Low airflow Alarm and Supply & Exhaust filter Fail |